0

Total Items 0

  • Email: info@cem-instruments.in
  • Phone: +91-33-22151376 / 22159759

Dry Well Temperature Calibrator


Dry-Well Temperature

BX-150

The new heat source to check your thermometer’s accuracy up to 300°C/572°F. Simply set the heat source to the desired test point, insert the test probe into the best fitting well, and get a verification of the performance. It’s that easy!

add to enquiry Download PDF

General Specifications

Temperature Range +33  to 300°C (at 20°C ambient )
Accuracy 0.8°C; Temp 100°C
(±1.6°F; Temp 212°F)
±1.6°C; 100°C Temp 200°C
(±3.2°F; 212°F Temp 392°F)
±2.8°C; 200°C Temp 300°C
(±5.6°F; 392°F Temp 572°F)
Stability ±0.1°C; Temp 100°C
(±0.2°F; Temp 212°F)
±0.2°C; 100°C Temp 200°C
(±0.4°F; 212°F Temp 392°F)
±0.4°C; 200°C Temp 300°C
(±0.8°F; 392°F Temp 572°F)
Resolution 0.1 °C/°F
Heating Time 30  minutes to max
Cooling Time 30  minutes to 100°C (122°F )
Aperture Diameter Φ 4.2mm  Φ 6.8mm  Φ 5mm  Φ 4.5 mm
Power 230 VAC (±10%), 1.25A; or 110VAC (±10%), 2.5A
Size (H*W*D) 180 mm*114mm*233mm
Weight 3 kg

 

ENVIRONMENTAL CONDITIONS

  • The instrument operates safely under the follwing conditions:
  • Temperature Range : 5-35°C (41-95°F)
  • Ambient Relative Humidity : 15-80%
  • Pressure : 75kPa - 106kPa
  • Mains Voltage within ±10% of nominal
  • Vibrations in the calibration Environment should be Minimized

 

 

 

  • BX 150 dry-block heat sources
  • Power Cord
  • User’s Guide
  • Calibration Report